Calibrate Perkin Elmer UV/VIS Lambda 900
Wavelength calibration
run program c:\uvwinlab\service9.exe
The spectrometer needs to be switched on for at least an hour (indicated in the program).
Also make sure the white light is visible on moving the wavelength to 0 nm. Otherwise make a scan from 2000 → 400 nm. This usually solves it.
It calibrates :
- Slit
- UV/VIS 2 peaks (using the 656.1 nm from the internal deuterium lamp, and it's zero order)
- NIR 2 peaks (using the 2nd order of the 656.1 nm)
- 0% Transmission
This takes about 10-15 minutes, after which “Calibrated successfully” is displayed.